Warranty for the whole machine is 1 year, for the motor is 2 years, and for the battery and other fast consumption items is 3 months.
Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ), if the temperature is over 135 °C, the motor will be damaged.
Normal bone drill can only hold drill bits less than 6mm, acetabulum reaming drill with big torque, rpm is around 300, canulate drill is used with nail or K wire, rpm is around 600.
The battery needs to be taken out of the power tool after surgery. Otherwise, the small automatic discharge of the electric will make the life of the battery short.
The battery cannot be sterilized, it is not autoclavable.
The cranial bur must be washed right after the surgery by purger in order to avoid the blood sticking in the bur, then we should make it dry.
The battery space in the handpiece of the tool should not have any water after sterilization.
The cranial bur blade is made of medical stainless steel, in order to avoid breaking, so it is flexible and not very hard, it is easy to be attrite, we suggest each blade only be used for 1 surgery.
Packing: Optional Aluminum case for each set, then into a good quality carton, or on request.
Shipping: Door to door courier like DHL/TNT/UPS/FEDEX, or shipped by air/sea.
Frequently Asked Questions (FAQ)
What is the oscillation speed of this orthopedic small oscillating saw?
The oscillation speed range is 0-16000 cpm, featuring no-limit speed-changing for precise control during bone cutting procedures.
Can the battery pack be sterilized in an autoclave?
No, the battery pack cannot be sterilized and is not autoclavable. Please ensure the battery is removed before autoclaving the handpiece.
What is the maximum autoclave temperature allowed for the handpiece?
The standard sterilizing temperature should be kept at or below 135 °C. For the improved type, it can go up to 155 °C. Exceeding these limits may cause permanent damage to the motor.
Why should I remove the battery immediately after surgery?
Leaving the battery in the tool after surgery allows automatic micro-discharge, which can significantly shorten the battery lifespan. Always store them separately.
What is the warranty period for this medical saw system?
We offer a 1-year warranty for the complete machine body, 2 years for the integrated motor, and 3 months for consumables including the battery.
How often should the cranial bur blades be replaced?
The cranial bur blades are made of medical stainless steel, which is flexible to prevent breakages but will wear out easily. We highly recommend using a fresh blade for each individual surgery.